Installing a Reagent Bottle or Bag in the Refrigerator
This tutorial contains two methods for installing the reagent bottle or bag in the Refrigerator:
Important
Wear appropriate PPE.
Required Items:
1 pack - Formulatrix customized cap (max. 12x autoclave cycles per cap assembly)
A Package of Formulatrix Customized Cap
1 pack - Sterilized pinch valve tube
A Package of Pinch Valve Tube
1 ea - Hydrophobic filter
A Hydrophobic Filter
1 ea - Autoclaved female luer plug. This plug will secure the old bottle dispensing tube during replacement.
A Female Luer Plug
2 ea - Autoclaved male luer plug. This plug will secure the tubes during replacement.
A Male Luer Plug
1 ea - Reagent bottle or bag
1 ea - 70% Ethanol in a spray bottle
1 ea - Falcon tube containing 70% ethanol or 70% IPA
1 ea - Falcon tube containing sterile water
1 ea - Kimwipes or similar delicate task wipers
To Install a Reagent Bottle or Bag Cap/Tubing in the Refrigerator
Section 1: Replacing the Formulatrix Customized Reagent Bottle Cap
Note
If you use a reagent bag/bottle with a Formulatrix customized reagent bottle cap, skip to Section 2: Opening Refrigerator.
Using the spray bottle, apply 70% ethanol solution to the reagent bottle, the hydrophobic filter package, and the Formulatrix customized cap package in order to ensure they are clean and sterile.
Place the treated items in the biosafety cabinet (BSC).
Tighten the bottle dispensing and vent tube clamps until you hear a clicking sound.
Replace the original cap of the reagent bottle with the Formulatrix customized cap.
Important
Make sure the vent tube end inside the bottle is not blocked, allowing air to pass through freely.
Connect the other end of the vent tube to the hydrophobic filter’s outlet.
![]()
A Formulatrix Customized Cap with a Hydrophobic Filter Installed
Section 2: Opening Refrigerator
Press the Pause/Resume button below the door handles to pause active tasks. The button’s light should blink blue while active tasks are paused.
Once the pause/resume button turns solid blue, open the refrigerator door.
Section 3: Installing a New Pinch Valve Tube and Reagent Bottle/Bag
Note
Skip to Section 4: Refilling/Swapping the Reagent Bottle/Bag if you are refilling/swapping the old reagent bottle/bag.
3A: Installing New Pinch Valve Tube
Open the Refrigerator.
Place the new valve tube package into the media storage area.
Note
If you are using a new (and autoclaved) valve tube, use the sprayer to apply 70% ethanol before installation.
Ensure all pinch clamps are tight.
Remove the silicone hole cover at the top of the dispensing area that matches the hanger’s number, and insert the pinch valve tube. Replace the silicone hole cover with the tube in the hole.
Guide the end of the pinch valve tube with the covered dispensing nozzle through the hole to the dispensing area.
Note
If you reuse the dispensing nozzle, make sure it has been autoclaved and covered with aluminum foil.
Once the pinch valve tube of sufficient length is in the dispensing area, open the window.
Press the Valve button on the dispensing station. It should turn green and you should hear a clicking sound. The valve will open for 10 seconds.
Important
The hanger, hole, and valve numbers must match.
Insert the pinch valve tube into the pinch valve in the dispensing station.
Important
Hold the side of the dispenser nozzle when installing it on the pinch valve tube. Avoid touching the dispenser nozzle’s end to prevent contamination.
Pull the tube up until the dispenser nozzle is underneath the valve.
![]()
Installing the Dispenser Nozzle in the Valve
Press the Valve button again to fasten the valve if the installation is completed before the timeout. The green light should turn OFF.
Note
Use the spray bottle with 70% ethanol to treat the pinch valve tube and dispenser nozzle if necessary.
Close the dispensing area window.
Check the pinch valve tube for drooping and adjust if necessary.
Close the refrigerator door.
3B: Installing the Reagent Bottle/Bag
Open the Refrigerator.
Use the spray bottle to apply 70% ethanol to the reagent bottle or bag and the Formulatrix customized cap.
Place the treated items into the media storage area.
Prepare both Falcon tubes containing 70% ethanol and sterile water in the media storage area.
Note
If necessary, use a tube rack to hold the Falcon tubes. Use the sprayer with 70% ethanol to treat the rack prior to use.
Remove the luer plugs from the bottle dispensing tube and the pinch valve tube.
Dip the end of the bottle dispensing tube luer into sterile water to rinse residual reagent. Then, dip it into the 70% ethanol to disinfect.
Important
Make sure the ends of the tubes are dry. The 70% ethanol will quickly evaporate.
Connect the luer fittings of the bottle dispensing tube on the new bottle and the pinch valve tube.
Place the bottle on the hanger, connect the vent tube, and release all tube clamps. If necessary, use Kimwipes to clean and disinfect the media storage platform with 70% ethanol.
Important
If you are replacing reagent bottles and tubes, remove the old ones from the media storage area.
![]()
The Correct Positioning of Reagent Bottle and Tubes
Proceed to Section 5: Closing Refrigerator and Continuing Tasks.
Section 4: Refilling/Swapping the Reagent Bottle/Bag
A. Refilling The Bottle Or Bag With The Same Reagent
A1: Removing the Reagent Bottle/Bag and Tubes from Media Storage
Using the sprayer, treat your gloves and the autoclaved luer plugs package with 70% ethanol before accessing the media storage.
![]()
The Media Storage Area
Tighten all clamps until they click.
Remove the bottle from the hanger and place it in the media storage area.
Disconnect the bottle dispensing tube from the pinch valve tube at the bottle’s luer connection.
Install the autoclaved plugs on the bottle dispensing and pinch valve tubes to avoid leaking. Clean any spilled reagent with the Kimwipes.
Remove the reagent bottle from the Refrigerator and close the doors.
A2: Refilling The Reagent Bottle/Bag
Place the reagent bottle into the BSC.
Raise the bottle dispensing tube above the bottle and open the pinch clamp.
Note
Slightly loosen the autoclaved plug to draw the remaining reagent from the tube back into the bottle.
Tighten the pinch clamp and the autoclaved plug.
Open the Formulatrix customized cap and refill the reagent bottle.
Once refilled, securely close the reagent bottle and return it to the Refrigerator.
Go to section 3B: Installing the Reagent Bottle/Bag.
B. Swapping The Old Bottle/Bag Dispensing Tube With The Same Reagent
Remove the reagent bottle and tubes from media storage.
Install the new reagent bottle/bag.
C. Swapping The Old Bottle/Bag Dispensing Tube With a Different Reagent
C1: Removing The Old Bottle/Bag and Pinch Valve Tube
Remove the reagent bottle and tubes from media storage.
Open the Refrigerator.
Press and hold the handlebar to lift the dispensing area window.
![]()
The Dispensing Area
Clean the dispenser nozzle tip with a Kimwipe to remove reagent.
Remove the pinch valve tube from the pinch valve:
Press the Valve button on the dispensing station to open the pinch valve for 10 seconds. It should turn green and you should hear a clicking sound.
Important
The hanger, hole, and valve numbers must match.
![]()
The Valve Buttons on the Dispensing Station
Remove the tube within 10 seconds.
Note
If the pinch valve tightens before the tube is successfully removed, press the button again.
Once complete, close the dispensing area window.
Remove the silicone hole cover to remove the pinch valve tube from the dispensing area.
Press and hold the handlebar to lower the dispensing area window. Ensure the area is clean.
Reinstall the silicone hole cover after removing the pinch valve tube.
![]()
The Silicone Hole Cover
Remove the old reagent bottle and tubes from the media storage area and close the Refrigerator.
C2: Installing The New Pinch Valve Tube and Reagent Bottle/Bag
Install the new pinch valve tube.
Install the new reagent bottle/bag.
Section 5: Closing Refrigerator and Continuing Tasks
Close the refrigerator door. The UV lights will automatically turn ON for a few minutes to sterilize the clean chamber area whenever the refrigerator door is closed. A Caution! UVC Light indicator on the lower front part of the scale is illuminated when the UV lights are on.
![]()
The UV Lights in the Clean Chamber Area
Press the blue Pause/Resume button on the door to continue active tasks. The button and status light should turn off, indicating that the Refrigerator is now operating in the automated mode.
Section 6: Updating Dispenser Reservoir in Cellmatic Suite
Open the Cellmatic Suite software and go to Inventory > Dispenser Reservoir.
For refilling or swapping bottle/bag with the same reagent: Select the Refill button
.
For swapping a bottle/bag with a different reagent:
Locate the hanger number where the different media bottle or bag is swapped.
Check the box next to I already removed the dispenser by following the correct procedures in the confirmation message.
Select OK to confirm.
![]()
The Confirmation Message to Remove Dispenser Reagent
Once the content is deleted, select the Plus sign to add information for the new dispenser reservoir.
On the Add New Reagent page, select the Reagent and fill out the Total Volume, Expiration Date, Reagent Shelf Life, and Minimal Volume Reminder. All fields with a red asterisk ( * ) are mandatory.
Mark I already installed the media bag by following the correct procedures under the User Confirmation.
Note
When a reagent expires, a red caution symbol will appear on the Consumables tab of the Dashboard page. Additionally, the expired reagent in the Dispenser Reagent column will be highlighted in red.
Select SAVE. The reagent should appear on the Dispenser Reservoir page.
![]()
The Add New Dispenser Reservoir Page