You are here: Chip Specifications Reference

Chip Specifications Reference

MANTIS CV and Volume Accuracy

For instructions to test MANTIS chip dispense volumes, see Testing MANTIS Chip Dispense Accuracy.

MANTIS Chip Specifications

Chip Type

Default Prime Value Default Pre-dispense Value Diaphragm Volume Minimum Volume Maximum Volume Pulses per second Dead Volume Accuracy Error Spec (LD) Accuracy Error Spec (SD) CV Spec Related Chip Lifetime (cycles)
Tube Pipette Tip Tube Pipette Tip
Low Volume (LV) Silicone Chip 100.4 µL 6 µL 0.6 µL 0.6 µL 0.1 µL, 0.5 µL 0.1 µL

At 0.1 µL= 10,
At 0.5 µL = 7

6 µL 10% 10% <3% 1,000,000
High Volume (HV) Silicone Chip 264 µL 16 µL 6 µL 6 µL 1 µL, 5 µL 1 µL At 1 µL= 10,
At 5 µL = 4
15 µL 10% 10% <3% 1,000,000
Low Volume PFE (LVP) Chip 100.4 µL 6 µL 0.6 µL 0.6 µL 0.1 µL, 0.5 µL 0.1 µL At 0.1 µL= 5,
At 0.5 µL = 4
6 µL 5% 10% <5% 750,000
High Volume PFE (HVP) Chip 264 µL 16 µL 6 µL 6 µL 1 µL, 5 µL 1 µL At 1 µL= 5,
At 5 µL = 4
15 µL 3% 8% <3% 750,000
Continuous Flow (CF) Silicone Chip 250 µL N/A N/A N/A N/A As low as ~5 µL Up to ~600 µL/sec ~150 µL N/A N/A <3% 1,000,000
Continuous Flow (CF) PFE Chip 250 µL N/A N/A N/A N/A As low as ~5 µL Up to ~600 µL/sec ~150 µL N/A N/A <3% 1,000,000

Fluid Channel Dimensions:

Material Composition:

Nozzle Selection:

 