Continuous Flow (CF) and High Viscosity Continuous Flow (HVCF) chips must be calibrated before you use them with new reagents. Chip calibration determines the amount of fluid dispensed per unit time.
In this section, you will learn how to calibrate a new reagent, use the example calibration data, or even copy any existing calibration data from a similar reagent.
See the table below to help select the appropriate continuous flow chips.
Description | Standard CF | HVCF |
---|---|---|
Minimum dispense volume |
~5 µL |
~10 µL |
Reagent Viscosity Range |
1 cP to 25 cP |
25 cP to 900 cP |
Media Bottle |
50 mL Falcon tubes, 250 mL Nalgene, GL45 capped glassware |
GL45 capped glassware |
Chip Nozzle Size for Calibration |
0.1 mm - 0.6 mm |
0.4 mm, 0.5 mm, 0.6 mm, 1 mm |
Tip: For quicker calibration, you can import a .CSV file of Continuous Flow calibration data from your computer to the MANTIS software. You can also export the calibration data of your calibrated reagents for future reference. For more information, see Exporting and Importing Continuous Flow Data Across MANTISes.
MANTIS software version 5.1 or newer
Continuous Flow feature license
MANTIS instrument V4
Note: The standard CF chip is also compatible with MANTIS instrument V3.
Either one of the following options:
The FORMULATRIX Calibration Kit with 6 microcentrifuge tubes. If you do not have the calibration kit, please contact support@formulatrix.com.
OR
Six weigh boats
A standard CF or HVCF chip and tubing harness
You have prepared the reagent to be calibrated (or a good surrogate with closely similar density, viscosity, and surface tension).
Note: To enable the Copy Existing Calibration Data option, you will need at least one Continuous Flow calibration data in the MANTIS database. Otherwise, this option will be hidden.
The Copy Existing Calibration Data Option is Active
Choose from the following options to open the Continuous Flow Calibration wizard:
Click the Input button of the uncalibrated reagent in the Input Control Panel. Select Continuous Flow or High Viscosity Continuous Flow from the Chip Type option. Then, click Start New Calibration.
Input Options Window
Note: If you calibrate the chip from the Input Options window, MANTIS automatically defines chip port position, chip serial number, and reagent name.
OR
Go to the Tools menu, and click Continuous Flow Options. Select Continuous Flow Calibration, and then click Next.
Continuous Flow Option Window
Select the calibration type from the following options:
Click New Calibration to calibrate a reagent for the first time, and click Next.
Click New Calibration
Select the chip port position of the assigned reagent. The chip’s serial number will automatically appear in the Chip Serial Number field. Click Next.
Select the Chip Port Position
Specify the calibration setup by following the steps below.
Important: For HVCF chip calibration, follow the HVCF Chip Calibration Recommendation for accurate results.
Next to Reagent Name, select the reagent you want to calibrate.
Select the Reagent Viscosity for the standard CF chip from 1 to 30 cP. This viscosity value is used for nozzle recommendation and does not impact the calibrated volume.
Note: If you calibrate the HVCF chip, the Reagent Viscosity will be automatically generated to High Viscosity.
If you know the density of your reagent, select I know the Reagent Density and type the Density in g/mL. If you do not have the reagent density value, uncheck the I know the Reagent Density checkbox.
Select the Nozzle Size from the available options.
Select the appropriate Bottle Pressure value from the following options: 3, 6, 10, 13, and 16 psi.
Select the Bottle Size accordingly: Small (100 mL), Medium (100-250 mL), or Large (250-500 mL). Recommended Bottle Size for HVCF chip is Medium (100-250 mL).
Set the Calibration Data for the Selected Reagent
Important: If this warning indicator appears next to the Nozzle Size or Bottle Pressure field, hover over it to see the recommended bottle pressure and nozzle size values from MANTIS.
Warning Indicators
Click Yes, I have the Calibration Kit if you want to use a tube rack and Eppendorf-style tubes for calibrating. Then, click Next.
OR
Click No, I don’t have the Calibration Kit if you want to use weigh boats. Then, click Next.
Specify the Calibration Kit
The table below explains in detail the required preparation for the microcentrifuge tubes and the weigh boats.
Calibration Tools | Instructions | ||||
---|---|---|---|---|---|
Microcentrifuge Tubes |
|
||||
Weigh Boats |
|
Select Copy Existing Calibration Data, and then click Next.
Select Copy Existing Calibration Data
Specify the calibration setup by following the steps below.
Define the Reagent Name. If you copy calibration data from the Input Options window, the reagent name will be automatically generated.
Select the desired calibration data from the Select Existing Calibration Data list. The copied calibration data will be automatically displayed in each respective field.
Click Copy Calibration.
The Existing Calibration Data
Click Finish.
Select this option to calibrate the CF chip for use with any reagent with approximately 1 cP viscosity (water-like).
Important: This option is unavailable for HVCF chip calibration.
Select Use Example Calibration, then click Next.
At this stage, the Reagent Viscosity field has been automatically assigned to 1 cP.
Define the nozzle size, bottle pressure, and bottle size in the provided fields.
Use Example Data for 1 cP Viscosity Reagent
Click Apply Calibration to apply the setup value.
Click Finish.
Calibration completed! The calibration results will be generated in the last page of the wizard. Click Finish.
The Calibration Result
Note: If you calibrate a CF chip from the Input Options window, the calibration data will automatically be generated under the Calibration Data section. Click OK to save the changes.
The CF Calibration Data
The following tables are for general reference. The calibration results may vary depending on the chip lifetime, nozzle size, and pressure used during calibration.
1 mm Nozzle | ||||||
---|---|---|---|---|---|---|
Glycerin Mix | Viscosity (cP) | Pressure Settings | ||||
3 Psi | 6 Psi | 10 Psi | 13 Psi | 16 Psi | ||
Distilled Water |
1 | - | - | - | - | - |
70% |
25 | ✔ | - | - | - | - |
50 | ✔ | ✔ | ✔ | ✔ | ✔ | |
80% |
100 | ✔ | ✔ | ✔ | ✔ | ✔ |
150 | ✔ | ✔ | ✔ | ✔ | ✔ | |
200 | ✔ | ✔ | ✔ | ✔ | ✔ | |
250 | ✔ | ✔ | ✔ | ✔ | ✔ | |
300 | ✔ | ✔ | ✔ | ✔ | ✔ | |
350 | - | ✔ | ✔ | ✔ | ✔ | |
90% |
400 | - | ✔ | ✔ | ✔ | ✔ |
450 | - | ✔ | ✔ | ✔ | ✔ | |
500 | - | ✔ | ✔ | ✔ | ✔ | |
550 | - | ✔ | ✔ | ✔ | ✔ | |
600 | - | ✔ | ✔ | ✔ | ✔ | |
650 | - | ✔ | ✔ | ✔ | ✔ | |
100% | 700 | - | - | - | ✔ | ✔ |
750 | - | - | - | ✔ | ✔ | |
800 | - | - | - | ✔ | ✔ | |
850 | - | - | - | ✔ | ✔ | |
900 | - | - | - | ✔ | ✔ |
0.6 mm Nozzle | ||||||
---|---|---|---|---|---|---|
Glycerin Mix | Viscosity (cP) | Pressure Settings | ||||
3 Psi | 6 Psi | 10 Psi | 13 Psi | 16 Psi | ||
Distilled Water |
1 | - | - | - | - | - |
70% |
25 | ✔ | - | - | - | - |
50 | ✔ | ✔ | ✔ | ✔ | ✔ | |
80% |
100 | - | ✔ | ✔ | ✔ | ✔ |
150 | - | ✔ | ✔ | ✔ | ✔ | |
200 | - | ✔ | ✔ | ✔ | ✔ | |
250 | - | ✔ | ✔ | ✔ | ✔ | |
300 | - | ✔ | ✔ | ✔ | ✔ | |
350 | - | ✔ | ✔ | ✔ | ✔ | |
90% |
400 | - | - | ✔ | ✔ | ✔ |
450 | - | - | ✔ | ✔ | ✔ | |
500 | - | - | ✔ | ✔ | ✔ | |
550 | - | - | ✔ | ✔ | ✔ | |
600 | - | - | ✔ | ✔ | ✔ | |
650 | - | - | ✔ | ✔ | ✔ | |
100% | 700 | - | - | - | ✔ | ✔ |
750 | - | - | - | ✔ | ✔ | |
800 | - | - | - | ✔ | ✔ | |
850 | - | - | - | ✔ | ✔ | |
900 | - | - | - | ✔ | ✔ |
0.5 mm Nozzle | ||||||
---|---|---|---|---|---|---|
Glycerin Mix | Viscosity (cP) | Pressure Settings | ||||
3 Psi | 6 Psi | 10 Psi | 13 Psi | 16 Psi | ||
Distilled Water |
1 | - | - | - | - | - |
70% |
25 | ✔ | - | - | - | - |
50 | ✔ | ✔ | ✔ | ✔ | ✔ | |
80% |
100 | - | ✔ | ✔ | ✔ | ✔ |
150 | - | ✔ | ✔ | ✔ | ✔ | |
200 | - | ✔ | ✔ | ✔ | ✔ | |
250 | - | ✔ | ✔ | ✔ | ✔ | |
300 | - | ✔ | ✔ | ✔ | ✔ | |
350 | - | ✔ | ✔ | ✔ | ✔ | |
90% |
400 | - | - | - | ✔ | ✔ |
450 | - | - | - | ✔ | ✔ | |
500 | - | - | - | ✔ | ✔ | |
550 | - | - | - | ✔ | ✔ | |
600 | - | - | - | ✔ | ✔ | |
650 | - | - | - | ✔ | ✔ | |
100% | 700 | - | - | - | - | - |
750 | - | - | - | - | - | |
800 | - | - | - | - | - | |
850 | - | - | - | - | - | |
900 | - | - | - | - | - |
0.4 mm Nozzle | ||||||
---|---|---|---|---|---|---|
Glycerin Mix | Viscosity (cP) | Pressure Settings | ||||
3 Psi | 6 Psi | 10 Psi | 13 Psi | 16 Psi | ||
Distilled Water |
1 | - | - | - | - | - |
70% |
25 | ✔ | - | - | - | - |
50 | ✔ | ✔ | ✔ | ✔ | ✔ | |
80% |
100 | - | ✔ | ✔ | ✔ | ✔ |
150 | - | ✔ | ✔ | ✔ | ✔ | |
200 | - | ✔ | ✔ | ✔ | ✔ | |
250 | - | ✔ | ✔ | ✔ | ✔ | |
300 | - | ✔ | ✔ | ✔ | ✔ | |
350 | - | ✔ | ✔ | ✔ | ✔ | |
90% |
400 | - | - | - | ✔ | ✔ |
450 | - | - | - | ✔ | ✔ | |
500 | - | - | - | ✔ | ✔ | |
550 | - | - | - | ✔ | ✔ | |
600 | - | - | - | ✔ | ✔ | |
650 | - | - | - | ✔ | ✔ | |
100% | 700 | - | - | - | - | - |
750 | - | - | - | - | - | |
800 | - | - | - | - | - | |
850 | - | - | - | - | - | |
900 | - | - | - | - | - |
Important: For the most accurate dispense possible, we recommend recalibrating your reagents each time you change CF chips with New Calibration option. Follow the instructions of the New Calibration above. The previous calibration data will be automatically replaced by the new calibration setup.
![]() |
|
MAOWH-V5R124 |