FORMULATRIX offers a variety of chip types for a wide range of volumes and chemical compatibility to achieve the desired Mantis dispense performance. If you wish to purchase the Mantis chips and chip nozzles, please visit the Mantis web shop or contact FORMULATRIX sales representative at sales@formulatrix.com. .

Here is the list of Mantis chip types, chip specifications, and the available chip nozzle sizes.

Chip Types and Specifications

Chip Type Default Prime Value Default Pre-dispense Value Diaphragm Volume Minimum Volume Maximum Volume Pulses per second Hold-up Volume Dead Volume Accuracy Error Spec (LD) Accuracy Error Spec (SD) CV Spec Rated Chip Lifetime (cycles)
Tube Pipette Tip Tube Pipette Tip LD SD

Silicone Low Volume (LV)

100.4 µL 6 µL 0.6 µL 0.6 µL 0.5 µL 0.1 µL 0.1 µL ∞

At 0.1 µL= 10, At 0.5 µL = 7

60 µL Default 3.6 µL** 10% 10% <3% 1,000,000

Silicone High Volume (HV)

264 µL 16 µL 6 µL 6 µL 5 µL 1 µL 1 µL ∞ At 1 µL= 10, At 5 µL = 4 200 µL Default 36 µL** 10% 10% <3% 1,000,000

Low Volume 3PFE (LVP)

100.4 µL 6 µL 0.6 µL 0.6 µL 0.5 µL 0.1 µL 0.1 µL ∞ At 0.1 µL= 3, At 0.5 µL = 2 60 µL Default 3.6 µL** 5% 8% <3% 750,000

High Volume 3PFE (HVP)

264 µL 16 µL 6 µL 6 µL 5 µL 1 µL 1 µL ∞ At 1 µL= 6, At 5 µL = 4 200 µL Default 36 µL** 5% 8% <3% 750,000

Silicone Low Volume (LV) Hi-Fi*

100.4 µL 6 µL 0.6 µL 0.6 µL 0.5 µL 0.1 µL 0.1 µL ∞ At 0.1 µL= 10, At 0.5 µL = 7 60 µL Default 3.6 µL** 5% 5% <2% 1,000,000

Silicone High Volume (HV) Hi-Fi*

264 µL 16 µL 6 µL 6 µL 5 µL 1 µL 1 µL ∞ At 1 µL= 10, At 5 µL = 4 200 µL Default 36 µL** 5% 5% <2% 1,000,000

*Compatible with MANTIS 5.1 Software or newer

**User configurable

 

XL Volume PFE Chip

Chip Type Default Prime Value Default Pre-dispense Value Diaphragm Volume Minimum Volume Maximum Volume Pulses per second Hold-up Volume Dead Volume Accuracy Error Spec CV Spec Rated Chip Lifetime (cycles)
Tube Pipette Tip Tube Pipette Tip

XL Volume PFE (XLP)*

300 µL 150 µL 150 µL 50 µL 50 µL 50 µL ∞ At 50 µL/0.735 sec 230 µL Default 400 µL** 5% <2% 750,000

*Compatible for Mantis 5.2.1 Software or newer

**User configurable

 

Continuous Flow Chips

Chip Type Default Prime Value Default Pre-dispense Value Diaphragm Volume Minimum Volume Maximum Volume Pulses per second Hold-up Volume Dead Volume Accuracy Error Spec CV Spec Rated Chip Lifetime (cycles) Recovery Duration
Tube Pipette Tip Tube Pipette Tip

Silicone Continuous Flow (CF)

250 µL/ pulse N/A N/A N/A - As low as ~5 µL ∞ Up to ~600 µL/sec 350 µL 150 µL*** 3% <3% 1,000,000 500 ms

Continuous Flow (CF) PFE

250 µL/ pulse N/A N/A N/A - As low as ~5 µL ∞ Up to ~600 µL/sec 350 µL 150 µL*** 3% <3% 1,000,000 500 ms

Silicone High Viscous Continuous Flow (HVCF)**

250 µL/pulse N/A N/A N/A - As low as ~10 µL ∞ Up to ~130 µL/sec* 2800 µL 200 µL*** 5% <3% 1,000,000 500 ms

*using 100% glycerol or ±700 cP reagent

**Compatible for Mantis 5.1 Software or newer

***User configurable

 

Fluid Channel Dimensions:

The following information contains the diameter of the chip’s fluid channel at its narrowest point.

Chip Type Diameter
Low Volume (LV) 0.25 mm
High Volume (HV) 0.38 mm
Continuous Flow (CF) 0.8 mm
High Viscous Continuous Flow (HVCF) 1.1 mm
XL Volume PFE (XLP) 0.89 mm

Chip Tubing Dimensions:

The following information provides the inner and outer diameters of the chip tubing.

Chip Type Tubing Type Inner Diameter (mm) Outer Diameter (mm) Color
High Volume (HV) - 0.76 1.59 Green Translucent
High Volume High Fidelity (HV Hi-Fi)
High Volume PFE (HVP)
High Volume 3 PFE (HV3PFE)
Low Volume (LV) 0.4 1.59 Natural (Translucent)
Low Volume High Fidelity (LV Hi-Fi)
Low Volume PFE (LVP)
Low Volume 3 PFE (LV3PFE)
Continuous Flow (CF) Tubing Liquid 0.76 1.59 Natural (Translucent)
Tubing Air Pressure 1.02 2.16 Clear (silicone)
Continuous Flow (CF) PFE Tubing Liquid 0.76 1.59 Natural (Translucent)
Tubing Air Pressure 1.02 2.16 Clear (silicone)
High Viscous Continuous Flow (HVCF) Tubing Liquid 2.38 3.97 Clear
Tubing Air Pressure 1.02 2.16 Clear (silicone)
XL Volume PFE (XLP) - 0.76 1.59 Green Translucent

Material Composition:

Nozzle Selection:

For instructions to test Mantis chip dispense volumes, see Testing Mantis Chip Dispense Accuracy.

 

Chip’s Nozzle Kits Catalog

Note: All sizes of nozzle strip can be used with Low Volume, High Volume, and Continuous Flow chips.

Image and Nozzle Type Description
Color* Inner Diameter Size (mm) Part Number Notes
Red

0.1

232994  
Dark Brown 0.15 232289
  • For use with the Low Volume PFE Standard and CF PFE Standard chips.

  • Use this nozzle to calibrate reagents for use with the Continuous Flow (PFE) chip.

Blue 0.2 232995 For use with the Low Volume Standard chip.
Yellow 0.3 232996  
Dark Brown 0.3 232518 For use with the High Volume PFE Standard chip.
Peach 0.4 232997 Recommended to use with HVCF chip.
Green 0.5 206237
  • For use with the High Volume Standard chip.

  • Recommended to use with HVCF chip.

Maroon 0.6 232998 Recommended to use with HVCF chip.
Brown 0.8 836771A For use with the XL Volume PFE chip.
Translucent 1 819023A Recommended to use with HVCF chip.
Mantis Nozzle Kit - Silicone (2 each) Assorted 0.1, 0.2, 0.3, 0.4, 0.5, 0.6 233036  

*The color of the actual nozzles might be slightly different from the images displayed above.

Note: Should you experience any dispensing issues after changing nozzles, see Troubleshooting or contact support@formulatrix.com for further assistance.

 

Related Topics


MAOWH-V55R126

Formulatrix Trading FZCO

2E 104b, First Floor, 2 East

Dubai Airport Freezone

Dubai, UAE

 

+1-781-788-0228

support@formulatrix.com

publications@formulatrix.com