FORMULATRIX offers a variety of chip types for a wide range of volumes and chemical compatibility to achieve the desired Mantis dispense performance. If you wish to purchase the Mantis chips and chip nozzles, please visit the Mantis web shop or contact FORMULATRIX sales representative at sales@formulatrix.com. .
Here is the list of Mantis chip types, chip specifications, and the available chip nozzle sizes.
| Chip Type | Default Prime Value | Default Pre-dispense Value | Diaphragm Volume | Minimum Volume | Maximum Volume | Pulses per second | Hold-up Volume | Dead Volume | Accuracy Error Spec (LD) | Accuracy Error Spec (SD) | CV Spec | Rated Chip Lifetime (cycles) | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Tube | Pipette Tip | Tube | Pipette Tip | LD | SD | ||||||||||
|
Silicone Low Volume (LV)
|
100.4 µL | 6 µL | 0.6 µL | 0.6 µL | 0.5 µL | 0.1 µL | 0.1 µL | ∞ |
At 0.1 µL= 10, At 0.5 µL = 7 |
60 µL | Default 3.6 µL** | 10% | 10% | <3% | 1,000,000 |
|
Silicone High Volume (HV)
|
264 µL | 16 µL | 6 µL | 6 µL | 5 µL | 1 µL | 1 µL | ∞ | At 1 µL= 10, At 5 µL = 4 | 200 µL | Default 36 µL** | 10% | 10% | <3% | 1,000,000 |
|
Low Volume 3PFE (LVP)
|
100.4 µL | 6 µL | 0.6 µL | 0.6 µL | 0.5 µL | 0.1 µL | 0.1 µL | ∞ | At 0.1 µL= 3, At 0.5 µL = 2 | 60 µL | Default 3.6 µL** | 5% | 8% | <3% | 750,000 |
|
High Volume 3PFE (HVP)
|
264 µL | 16 µL | 6 µL | 6 µL | 5 µL | 1 µL | 1 µL | ∞ | At 1 µL= 6, At 5 µL = 4 | 200 µL | Default 36 µL** | 5% | 8% | <3% | 750,000 |
|
Silicone Low Volume (LV) Hi-Fi*
|
100.4 µL | 6 µL | 0.6 µL | 0.6 µL | 0.5 µL | 0.1 µL | 0.1 µL | ∞ | At 0.1 µL= 10, At 0.5 µL = 7 | 60 µL | Default 3.6 µL** | 5% | 5% | <2% | 1,000,000 |
|
Silicone High Volume (HV) Hi-Fi*
|
264 µL | 16 µL | 6 µL | 6 µL | 5 µL | 1 µL | 1 µL | ∞ | At 1 µL= 10, At 5 µL = 4 | 200 µL | Default 36 µL** | 5% | 5% | <2% | 1,000,000 |
*Compatible with MANTIS 5.1 Software or newer
**User configurable
XL Volume PFE Chip
| Chip Type | Default Prime Value | Default Pre-dispense Value | Diaphragm Volume | Minimum Volume | Maximum Volume | Pulses per second | Hold-up Volume | Dead Volume | Accuracy Error Spec | CV Spec | Rated Chip Lifetime (cycles) | ||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Tube | Pipette Tip | Tube | Pipette Tip | ||||||||||
|
XL Volume PFE (XLP)*
|
300 µL | 150 µL | 150 µL | 50 µL | 50 µL | 50 µL | ∞ | At 50 µL/0.735 sec | 230 µL | Default 400 µL** | 5% | <2% | 750,000 |
*Compatible for Mantis 5.2.1 Software or newer
**User configurable
Continuous Flow Chips
| Chip Type | Default Prime Value | Default Pre-dispense Value | Diaphragm Volume | Minimum Volume | Maximum Volume | Pulses per second | Hold-up Volume | Dead Volume | Accuracy Error Spec | CV Spec | Rated Chip Lifetime (cycles) | Recovery Duration | ||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Tube | Pipette Tip | Tube | Pipette Tip | |||||||||||
|
Silicone Continuous Flow (CF)
|
250 µL/ pulse | N/A | N/A | N/A | - | As low as ~5 µL | ∞ | Up to ~600 µL/sec | 350 µL | 150 µL*** | 3% | <3% | 1,000,000 | 500 ms |
|
Continuous Flow (CF) PFE
|
250 µL/ pulse | N/A | N/A | N/A | - | As low as ~5 µL | ∞ | Up to ~600 µL/sec | 350 µL | 150 µL*** | 3% | <3% | 1,000,000 | 500 ms |
|
Silicone High Viscous Continuous Flow (HVCF)**
|
250 µL/pulse | N/A | N/A | N/A | - | As low as ~10 µL | ∞ | Up to ~130 µL/sec* | 2800 µL | 200 µL*** | 5% | <3% | 1,000,000 | 500 ms |
*using 100% glycerol or ±700 cP reagent
**Compatible for Mantis 5.1 Software or newer
***User configurable
Fluid Channel Dimensions:
The following information contains the diameter of the chip’s fluid channel at its narrowest point.
| Chip Type | Diameter |
|---|---|
| Low Volume (LV) | 0.25 mm |
| High Volume (HV) | 0.38 mm |
| Continuous Flow (CF) | 0.8 mm |
| High Viscous Continuous Flow (HVCF) | 1.1 mm |
| XL Volume PFE (XLP) | 0.89 mm |
Chip Tubing Dimensions:
The following information provides the inner and outer diameters of the chip tubing.
| Chip Type | Tubing Type | Inner Diameter (mm) | Outer Diameter (mm) | Color |
|---|---|---|---|---|
| High Volume (HV) | - | 0.76 | 1.59 | Green Translucent |
| High Volume High Fidelity (HV Hi-Fi) | ||||
| High Volume PFE (HVP) | ||||
| High Volume 3 PFE (HV3PFE) | ||||
| Low Volume (LV) | 0.4 | 1.59 | Natural (Translucent) | |
| Low Volume High Fidelity (LV Hi-Fi) | ||||
| Low Volume PFE (LVP) | ||||
| Low Volume 3 PFE (LV3PFE) | ||||
| Continuous Flow (CF) | Tubing Liquid | 0.76 | 1.59 | Natural (Translucent) |
| Tubing Air Pressure | 1.02 | 2.16 | Clear (silicone) | |
| Continuous Flow (CF) PFE | Tubing Liquid | 0.76 | 1.59 | Natural (Translucent) |
| Tubing Air Pressure | 1.02 | 2.16 | Clear (silicone) | |
| High Viscous Continuous Flow (HVCF) | Tubing Liquid | 2.38 | 3.97 | Clear |
| Tubing Air Pressure | 1.02 | 2.16 | Clear (silicone) | |
| XL Volume PFE (XLP) | - | 0.76 | 1.59 | Green Translucent |
Material Composition:
Nozzle Selection:
For instructions to test Mantis chip dispense volumes, see Testing Mantis Chip Dispense Accuracy.
Note: All sizes of nozzle strip can be used with Low Volume, High Volume, and Continuous Flow chips.
| Image and Nozzle Type | Description | |||
|---|---|---|---|---|
| Color* | Inner Diameter Size (mm) | Part Number | Notes | |
|
Red |
0.1 |
232994 | |
|
Dark Brown | 0.15 | 232289 |
|
|
Blue | 0.2 | 232995 | For use with the Low Volume Standard chip. |
|
Yellow | 0.3 | 232996 | |
|
Dark Brown | 0.3 | 232518 | For use with the High Volume PFE Standard chip. |
|
Peach | 0.4 | 232997 | Recommended to use with HVCF chip. |
|
Green | 0.5 | 206237 |
|
|
Maroon | 0.6 | 232998 | Recommended to use with HVCF chip. |
|
Brown | 0.8 | 836771A | For use with the XL Volume PFE chip. |
|
Translucent | 1 | 819023A | Recommended to use with HVCF chip. |
| Mantis Nozzle Kit - Silicone (2 each) | Assorted | 0.1, 0.2, 0.3, 0.4, 0.5, 0.6 | 233036 | |
*The color of the actual nozzles might be slightly different from the images displayed above.
Note: Should you experience any dispensing issues after changing nozzles, see Troubleshooting or contact support@formulatrix.com for further assistance.
|
|
| MAOWH-V55R126 |