In the EXPLORER menu, locate the saved profile that you want to run.
Tap the RUN button located in the upper-right corner of your screen.
Then, wait for the Components Availability Checker to ensure that all required components are properly attached.
PULSE Components Availability Checker
In the Starting Volume Confirmation, tap CONFIRM if the sample and buffer volumes in the profile are the same as the actual volume in both of the tubes.
OR
If you are using different sample and buffer volumes, type the actual volumes and tap CONFIRM.
Starting Volume Confirmation Options
The profile will run and you can see the concentration progress in the STATUS menu, even from your external device. See Monitoring the PULSE Status from an External Device for more information.
Important: If you hear a hissing sound and an error message appears, the silicone seals are not properly aligned with the filter chip, which will result in leaking. Before continuing, check the chip and the silicone seals and reposition them if necessary, then try again. If this problem persists, please contact support@formulatrix.com for assistance. For more information, see Fixing Silicone Seals and Chip Leakage.
The cycle will pause and switch to Manual Control when the volume has reached the threshold (set from More Options) above your final volume. This is to allow you to check the volume before finishing the profile. From the STATUS menu, under the profile name, select one of the SWITCH TO MANUAL options below.
In the DEFAULT TIME, select one of the options in the drop down menu and tap CONTINUE.
Type the desired time in the CUSTOM TIME and tap CONTINUE.
Press and hold the HOLD TO RUN option to run the cycle and release it when you have finished.
Manual Control Options
Once your sample has reached the desired volume, tap FINISH and select one of the liquid recovery options.
NONE: Tap to skip the liquid recovery.
AIR: Tap to use air for the liquid recovery. Make sure your liquid level in the tubes is below the short tubing (<15 mL) so that PULSE can use the air.
BUFFER: Tap to use buffer for the liquid recovery.
AIR & BUFFER: Tap to use both for the liquid recovery.
Recovery of Liquid Options
When the cycle is complete, open the handle clamp and carefully remove the chip and tubes.
After running a profile, you may need to clean the tubing or the filter chips. See Cleaning Chips and Tubing for more information.
![]() |
|
PTFF-V15R020 |