You are here: Chip Specifications Reference

Chip Specifications Reference

TEMPEST CV and Volume Accuracy

For instructions to test TEMPEST chip dispense volumes, see Testing TEMPEST Chip Dispense Accuracy.

TEMPEST Chip Specifications

Chip Type Default Prime Value Default
Pre-dispense Value
Diaphragm Volume Minimum Volume Maximum Volume Pulses per second Dead Volume Accuracy Error Spec (LD) Accuracy Error Spec (SD) CV Spec Rated Chip Lifetime (cycles)
  Tube Pipette Tip Tube* Pipette Tip
Low Volume (LV) Silicone Chip 252 µL 70 µL 2.4 µL 0.2 µL, 1 µL 0.2 µL At 0.2 µL= 4,
At 1 µL = 3
50 µL 100 µL 8% 8% <5% at 0.2 µL
<5% at 1 µL
1,000,000
High Volume (HV) Silicone Chip 252 µL 70 µL 12.0 µL 1 µL, 5 µL 1 µL At 1 µL= 4,
At 5 µL = 3
50 µL 100 µL 5% 5% <4% at 1 µL
<4% at 5 µL
1,000,000
High Volume PFE (HVP) Chip 252 µL 70 µL 12.0 µL 1 µL, 5 µL 1 µL At 1 µL= 5,
At 5 µL = 3
50 µL 100 µL 10% 10% <5% 750,000

*Add 350 µL if not recoverable

Fluid Channel Dimensions:

Note:
1. See TEMPEST Chip Nozzle Strip Catalog for more information about TEMPEST chip’s nozzle strip type and size.
2. See the chip chemical compatibility chart by downloading it from the FORMULATRIX Micro-Diaphragm Chip Chemical Compatibility page.