For instructions to test TEMPEST chip dispense volumes, see Testing TEMPEST Chip Dispense Accuracy.
TEMPEST Chip Specifications
Chip Type | Default Prime Value | Default Pre-dispense Value |
Diaphragm Volume | Minimum Volume | Maximum Volume | Pulses per second | Dead Volume | Accuracy Error Spec (LD) | Accuracy Error Spec (SD) | CV Spec | Rated Chip Lifetime (cycles) | ||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Tube | Pipette Tip | Tube* | Pipette Tip | ||||||||||
Low Volume (LV) Silicone Chip | 510 µL | 70 µL | 2.4 µL | 0.2 µL, 1 µL | 0.2 µL | ∞ | At 0.2 µL= 4, At 1 µL = 3 |
50 µL | 100 µL | 8% | 8% | <5% at 0.2 µL <5% at 1 µL |
1,000,000 |
High Volume (HV) Silicone Chip | 400.8 µL | 70 µL | 12.0 µL | 1 µL, 5 µL | 1 µL | ∞ | At 1 µL= 4, At 5 µL = 3 |
50 µL | 100 µL | 5% | 5% | <4% at 1 µL
<4% at 5 µL |
1,000,000 |
High Volume PFE (HVP) Chip | 510 µL | 70 µL | 12.0 µL | 1 µL, 5 µL | 1 µL | ∞ | At 1 µL= 5, At 5 µL = 3 |
50 µL | 100 µL | 10% | 10% | <5% | 750,000 |
Continuous Flow (CF) Silicone Chip | 2000 µL | N/A | 50 µL | -- | As low as ~5 µL | ∞ | Up to ~600 µL/sec | ~150 µL | N/A | N/A | N/A | <3% | 1,000,000 |
*Add 350 µL if not recoverable
Fluid Channel Dimensions:
Material Composition:
Nozzle Selection:
TOW-V38R123
|