TEMPEST Chip Specifications

Chip Type Default Prime Value Default
Pre-dispense Value
Diaphragm Volume Minimum Volume Maximum Volume Pulses per second Dead Volume Accuracy Error Spec (LD) Accuracy Error Spec (SD) CV Spec Rated Chip Lifetime (cycles)
  Tube Pipette Tip Tube* Pipette Tip
Low Volume (LV) Silicone Chip 510 µL 70 µL 2.4 µL 0.2 µL, 1 µL 0.2 µL At 0.2 µL= 4,
At 1 µL = 3
50 µL 100 µL 8% 8% <5% at 0.2 µL
<5% at 1 µL
1,000,000
High Volume (HV) Silicone Chip 400.8 µL 70 µL 12.0 µL 1 µL, 5 µL 1 µL At 1 µL= 4,
At 5 µL = 3
50 µL 100 µL 5% 5% <4% at 1 µL
<4% at 5 µL
1,000,000
High Volume PFE (HVP) Chip 510 µL 70 µL 12.0 µL 1 µL, 5 µL 1 µL At 1 µL= 5,
At 5 µL = 3
50 µL 100 µL 10% 10% <5% 750,000
Continuous Flow (CF) Silicone Chip 2000 µL N/A 50 µL -- As low as ~5 µL Up to ~600 µL/sec ~150 µL N/A N/A N/A <3% 1,000,000

*Add 350 µL if not recoverable