You are here: Tutorials > Designing Profiles > Creating a Profile

Creating a Profile

µPULSE allows you to create a profile consisting of desired experiment settings for concentrating sample or performing buffer exchange. The system tracks volume as it runs sequences based on the values you have entered when designing a profile.

To set up a new profile, tap AUTO PROFILE on the Home Screen. Then choose between these 3 options:

 

AUTO PROFILE Menu

Tip: Tap the Help button to get acquainted with the software and learn about what you can do in the AUTO PROFILE menu.

 

Concentrate only

To create a new profile for concentration:

  1. On the Home Screen, open the AUTO PROFILE menu. The CONCENTRATE tab is selected by default.

  2. Type the Starting Volume in mL and Starting Concentration in mg/mL based on the sample you’re using, and then type the desired Final Volume. The Estimated Final Concentration will be automatically generated after you have set the other values.

  3. OPTIONAL: If you tapped the CONCENTRATE > 50 ML tab, type the volume in the Starting Volume of Tube 02 field. The default value in the 1st tube is 50 mL, but you can adjust it to your desired volume.

  4. Select the membrane cutoff size that matches the chip you’re using from the Membrane Cutoff Size drop-down list.

  5. OPTIONAL: Tap the Advanced Setup tab to configure advanced settings, such as pressure levels and pump timings, for each cycle. You can save the settings and load them for future use.

  6. Once you make sure everything is all set, tap RUN. See Running a Profile for more details.

 

Concentration + Buffer Exchange and Buffer Exchange Only

To create a new profile for concentration and buffer exchange:

  1. On the Home Screen, open the AUTO PROFILE menu and tap the BUFFER EXCHANGE tab.

  2. Type the Starting Volume in mL and Starting Concentration in mg/mL based on the sample you’re using, and then type the desired Final Volume. The Estimated Final Concentration will be automatically generated after you have set the other values.

    Note: When performing buffer exchange you cannot concentrate more than 50 mL.

  3. Then select between these two options:

  4. Type the desired value in the Sample Volume to Start Exchange field. This is the sample volume at which buffer exchange will begin.

  5. Then, type the exchange volume to use in the Exchange Volume field and input the total amount of buffer in the tube in the Total Buffer Volume field.

  6. Select the membrane cutoff size that matches the chip you’re using from the Membrane Cutoff Size drop-down list.

  7. OPTIONAL: Tap the Advanced Setup tab to configure advanced settings, such as pressure levels and pump timings, for each cycle. You can save the settings and load them for future use.

  8. Once you make sure everything is all set, tap RUN. See Running a Profile for more details.

     

Advanced Setup

You can configure more profile settings to meet your preferences by tapping the Advanced Setup tab. There you can find a list of advanced settings, such as pressure levels and pumping timings, for each cycle— i.e., prime, concentrate and buffer exchange, and recovery cycle.

µPULSE also allows you to save these settings for future use by tapping the SAVE SETUP AS button and type a specific name for it. The saved advanced setup that you are currently using for a profile is what will appear on the Advanced Setup Explorer tab on AUTO PROFILE menu. For example, if you use the system for the first time, DEFAULT SETUP appears as the name because the software uses the default advanced settings. 

Important: Make sure that you understand these options before making any changes.

Refer to the table below to better understand advance settings.

Note: Tap the RESET button to reset the values and switch back to the default values.

 

GENERAL

The Advanced Setup Explorer Tab and the General Sub-tab of Advanced Setup Tab

 

Items Description

Manual Control Threshold

This threshold sets the extra volume above the defined final volume. Once the threshold has been reached, the system pauses the concentration to avoid over-concentrating the sample.

Skip Priming Sequence

Set this option to skip priming with buffer from the right tube when µPULSE runs a profile.

 

MAIN CYCLE

Main Sub-tab of Advanced Setup

Items Description Default Value
Vacuum Pressure

Determines the vacuum pressure applied to the diaphragm.

-9.25 psi
Valve Pressure

Determines the pressure applied to the valves.

50 psi
Diaphragm Pressure

Determines the pressure applied to the diaphragm to make one pulse.

40 psi
Return Valve Pressure

Determines the resistance to flow that generates a trans-membrane pressure to drive concentration of the sample.

Note: Higher pressure results in faster concentration, but can damage the membrane if set too high.

Click the pencil icon to edit the value and select the membrane cutoff size if you haven’t already. The value you use here will only apply to the current advanced settings. Tap the reset button to reset this setting to the default Return Valve Pressure value.

Edit Return Valve Pressure Dialog Box

10 psi
Diaphragm Timing

Determines the timing for the diaphragm to make one full pulse.

275 ms
Valve Delay

Determines the operational timing of the valves.

25 ms

 

PRIME CYCLE

Prime Sub-tab of Advanced Setup

Items Description Default Value
Vacuum Pressure

Determines the vacuum pressure applied to the diaphragm.

-9 psi
Valve Pressure

Determines the pressure applied to the valves.

30 psi
Diaphragm Pressure

Determines the pressure applied to the diaphragm to make one pulse.

20 psi
Diaphragm Timing

Determines the timing for the diaphragm to make one full pulse.

500 ms
Valve Delay

Determines the operational timing of the valves.

50 ms

 

RECOVERY CYCLE

Recovery Sub-tab of Advanced Setup

Items Description Default Value
Vacuum Pressure

Determines the vacuum pressure applied to the diaphragm.

-9 psi
Valve Pressure

Determines the pressure value applied to the valves.

30 psi
Diaphragm Pressure

Determines the pressure applied to the diaphragm to make one pulse.

20 psi
Diaphragm Timing

Determines the timing for the diaphragm to make one full pulse.

500 ms
Valve Delay

Determines the operational timing of the valves.

50 ms

Related Topics


PUOWH-V21R022

10 DeAngelo Drive

Bedford, MA 01730

USA

 

+1-781-788-0228

support@formulatrix.com

publications@formulatrix.com