You are here: Chip Specifications Reference

Chip Specifications Reference

MANTIS CV and Volume Accuracy

For instructions to test MANTIS chip dispense volumes, see Testing MANTIS Chip Dispense Accuracy.

MANTIS Chip Specifications

Chip Type

Default Prime Value Default Pre-dispense Value Diaphragm Volume Minimum Volume Maximum Volume Pulses per second Dead Volume Accuracy Error Spec (LD) Accuracy Error Spec (SD) CV Spec Related Chip Lifetime (cycles)
Tube Pipette Tip Tube Pipette Tip
Low Volume (LV) Silicone Chip 100 µL 6 µL 0.6 µL 0.6 µL 0.1 µL, 0.5 µL 0.1 µL

At 0.1 µL= 10,
At 0.5 µL = 7

6 µL 10% 10% <3% 1,000,000
High Volume (HV) Silicone Chip 200 µL 16 µL 6 µL 6 µL 1 µL, 5 µL 1 µL At 1 µL= 10,
At 5 µL = 4
15 µL 10% 10% <3% 1,000,000
Low Volume PFE (LVP) Chip 100 µL 6 µL 0.6 µL 0.6 µL 0.1 µL, 0.5 µL 0.1 µL At 0.1 µL= 5,
At 0.5 µL = 4
6 µL 5% 10% <5% 750,000
High Volume PFE (HVP) Chip 200 µL 16 µL 6 µL 6 µL 1 µL, 5 µL 1 µL At 1 µL= 5,
At 5 µL = 4
15 µL 3% 8% <3% 750,000
Continuous Flow (CF) Silicone Chip 250 µL N/A N/A N/A N/A As low as ~5 µL Up to ~600 µL/sec ~150 µL N/A N/A <3% 1,000,000
Continuous Flow (CF) PFE Chip 250 µL N/A N/A N/A N/A As low as ~5 µL Up to ~600 µL/sec ~150 µL N/A N/A <3% 1,000,000

Fluid Channel Dimensions:

Note:
1. See MANTIS Chip's Nozzle Kits Catalog for more information about MANTIS chip’s nozzle type and size.
2. See the chip chemical compatibility chart by downloading it from the FORMULATRIX Micro-Diaphragm Chip Chemical Compatibility page.

Related Topics


MAOWH-V47R722

10 DeAngelo Drive

Bedford, MA 01730

USA

 

+1-781-788-0228

support@formulatrix.com

publications@formulatrix.com