For instructions to test MANTIS chip dispense volumes, see Testing MANTIS Chip Dispense Accuracy.
MANTIS Chip Specifications
Chip Type |
Default Prime Value | Default Pre-dispense Value | Diaphragm Volume | Minimum Volume | Maximum Volume | Pulses per second | Dead Volume | Accuracy Error Spec (LD) | Accuracy Error Spec (SD) | CV Spec | Related Chip Lifetime (cycles) | ||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Tube | Pipette Tip | Tube | Pipette Tip | ||||||||||
Low Volume (LV) Silicone Chip | 100 µL | 6 µL | 0.6 µL | 0.6 µL | 0.1 µL, 0.5 µL | 0.1 µL | ∞ |
At 0.1 µL= 10, |
6 µL | 10% | 10% | <3% | 1,000,000 |
High Volume (HV) Silicone Chip | 200 µL | 16 µL | 6 µL | 6 µL | 1 µL, 5 µL | 1 µL | ∞ | At 1 µL= 10, At 5 µL = 4 |
15 µL | 10% | 10% | <3% | 1,000,000 |
Low Volume PFE (LVP) Chip | 100 µL | 6 µL | 0.6 µL | 0.6 µL | 0.1 µL, 0.5 µL | 0.1 µL | ∞ | At 0.1 µL= 5,
At 0.5 µL = 4 |
6 µL | 5% | 10% | <5% | 750,000 |
High Volume PFE (HVP) Chip | 200 µL | 16 µL | 6 µL | 6 µL | 1 µL, 5 µL | 1 µL | ∞ | At 1 µL= 5,
At 5 µL = 4 |
15 µL | 3% | 8% | <3% | 750,000 |
Continuous Flow (CF) Silicone Chip | 250 µL | N/A | N/A | N/A | N/A | As low as ~5 µL | ∞ | Up to ~600 µL/sec | ~150 µL | N/A | N/A | <3% | 1,000,000 |
Continuous Flow (CF) PFE Chip | 250 µL | N/A | N/A | N/A | N/A | As low as ~5 µL | ∞ | Up to ~600 µL/sec | ~150 µL | N/A | N/A | <3% | 1,000,000 |
Fluid Channel Dimensions:
Note:
1. See MANTIS Chip's Nozzle Kits Catalog for more information about MANTIS chip’s nozzle type and size.
2. See the chip chemical compatibility chart by downloading it from the FORMULATRIX Micro-Diaphragm Chip Chemical Compatibility page.
MAOWH-V47R722 |