You are here: Maintenance : Testing MANTIS Chip Dispense Accuracy

Testing MANTIS Chip Dispense Accuracy

These tests should be performed to verify dispense volume accuracy and chip performance on the MANTIS automated liquid handler. These tests will identify dispense problems to assist in troubleshooting.

If you would like a full Operational Qualification (OQ) performed on your device, please contact the FORMULATRIX support team.

 

Dispense Volume Accuracy Test

There are two dispense volume accuracy tests:

  1. Dispense Volume Accuracy Test for HV Chip.
  2. Dispense Volume Accuracy Test for LV Chip.

Use the table below as a reference for the MANTIS chip specifications.

MANTIS Chip Type Accuracy Error Spec (LD) Accuracy Error Spec (SD) CV Spec Rated Chip Lifetime (cycles)
Low Volume (LV) Silicone Chip 10% 10% <3% 1,000,000
High Volume (HV) Silicone Chip 10% 10% <3% 1,000,000
Low Volume PFE (LVP) Chip 5% 10% <5% 750,000
High Volume PFE (HVP) Chip 3% 8% <3% 750,000
Low Volume PFE (LVP) Chip 5% 10% <5% 750,000
Continuous Flow (CF) Silicone Chip 3% N/A <3% 1,000,000
Continuous Flow (CF) PFE Chip 3% N/A <3% 1,000,000

Note: If your chip dispenses outside of the above specifications, repeat the tests using a new chip. Chips that are no longer dispensing accurately should be replaced. If multiple chips fail these tests, contact support@formulatrix.com for further assistance.

Required tools

 

Dispense Volume Accuracy Test for HV Chip

Dispense Volume Accuracy Test for LV Chip

 

Drop Location Accuracy and Chip Diaphragm Test

Required tools

Note: This test can be performed with a normal 96-well or 384-well plate, but it is recommended to use the clear reference plate, which is included in your MANTIS accessories kit. You can flip the reference plate upside-down to provide a flat, clear surface on which to dispense.

To test the drop location accuracy and chip diaphragm:

  1. In the MANTIS software, create a new dispense list and load the plate definition for the default 96-well or 384-well plate.

  2. Add your water reagent in the dispense list and assign it to the chip to be tested.

    Tip: Use water with food coloring to make drops easier to see.

  3. Set up your plate to dispense a single large diaphragm drop on one half of the plate, and a single small diaphragm drop on the other half. See Figure 9 below for an example of an HV chip (5 μL and 1 μL diaphragms) in a 96-well plate layout.

    Figure 9: Assign the 5 μL and 1 μL reagent volume in the Dispense List Designer in the 96-well plate

  4. Run the dispense onto the clear reference plate.

  5. Visually inspect the plate to check the following:

    1. The grid pattern should match the plate definition.

      Note: Drops may not be perfectly aligned due to static effects causing drops to shift on flat surfaces, but the grid should be mostly accurate. Major problems with drop placement may require service from the FORMULATRIX support team.

    2. The drops of each diaphragm should be consistent in size.

    3. There should be little or no satellite drops or splashing surrounding the drops.

      Note: If you see lots of small satellite drops, try replacing the nozzle on your chip.

  6. If your chip fails the above criteria, repeat this test using a new chip. If multiple chips fail this test, contact support@formulatrix.com for further assistance.

 

Related Topics


MAOWH-V47R722

10 DeAngelo Drive

Bedford, MA 01730

USA

 

+1-781-788-0228

support@formulatrix.com

publications@formulatrix.com