You are here: Software and Instrument Overview : Software Overview : Input Control Panel

Input Control Panel

The Input Control panel is represented by the image of the MANTIS at the bottom of the Device Control panel, where you can control the inputs as follows.

Input Control Panel Under the Device Control Panel

For more information, go to the respective topic in the Tutorials section.

 

Understanding the Input Control Panel

Use the image and table below to understand each of the components in the Input Control Panel.


Input Control Panel

Name Description
Input Toolbar The Input Toolbar provides you with eight dedicated buttons to perform specific MANTIS operations, explained below.
  • Auto Prime — Clicking the Auto Prime button primes the chip with a fixed amount of reagent that you can configure in the Input Option settings (see Prime Volume). Priming transfers reagent through the chip so that when you begin a dispense, the chip is not dispensing air. For more information, see Priming the Chips.

  • Manual Prime — Clicking the Manual Prime button will display the Manual Prime window. This option allows you to determine when to start and stop the priming process. For more information, see Priming the Chips.

    Manual Prime Window

  • Prime All Inputs — Clicking the Prime All button will prime every chip that has a reagent listed in the Reagent Input Field.

  • Wash Input — Clicking the Wash Input button will wash the chip currently stationed at the MANTIS hand.

  • Wash All Inputs — Clicking the Wash All Inputs button will wash every chip that has a reagent listed in the Reagent Input Field.

  • Wash Selected Inputs — Clicking the Wash Selected Inputs button will wash the chips of your preference that you have selected in the Wash Inputs Selection window. For more information, see Cleaning the Chips.

    Wash Inputs Selection Window

  • Recover Input — Clicking the Recover Input button runs the reagent recovery process on the chip currently stationed at the MANTIS hand. Recovering a reagent will return most of the liquid in the chip and tubing to the source container.

  • Recover All Inputs — Clicking the Recover All Inputs button runs the reagent recovery process on every chip that has a reagent listed in the Reagent Input Field. Recovering a reagent will return most of the liquid in the chip and tubing to the source container.

In addition to the above buttons, MANTIS also provides you with the two check bottle leak buttons, which only appear when Continuous Flow is active.

Check Bottle Leak Buttons on the Input Control Panel

  • Check Bottle Leak on Mantis Arm — Clicking the Check Bottle Leak on Mantis Arm button leads to the instrument checking for possible pressure leaks in the bottle attached to the chip on the MANTIS hand.

  • Check All Bottle Leak — Clicking the Check All Bottle Leak button checks all bottles for pressure leaks.

Device View Device View displays the MANTIS instrument in use. In the image displayed above, Chip 1 is on the MANTIS hand. Double-clicking one of the chips will cause that chip to be picked up by the MANTIS hand. Double-clicking a second time will return the chip from the MANTIS hand to the chip changer station. If you have the LC3 instrument, you will see the LC3 attached on the left or right side of the MANTIS.
Input The Input tells you which reagent is assigned to which chip station. To the right of each input field is a symbol or an abbreviation that represents the type of chip attached to that chip station. Options include:
  • LV — Low Volume Chip.

  • HV — High Volume Chip.

  • HVP — High Volume PFE Chip.

  • LVP — Low Volume PFE Chip.

  • HV3P — High Volume 3PFE Chip.

  • LV3P— Low Volume 3PFE Chip.

  • CF — Continuous Flow Chip.

You can see the status of your input through the Input Status box, whether it is defined in the Dispense List or not. This means that your input has been properly defined in the dispense list. You should check and define your reagent in the Dispense List if you find this image in your Input Status box.

To move the chips between inputs, you can drag the chips on the Input panel. The chip information will also be automatically updated.

Note: Before you drag the chip on the Input panel, make sure that you move the chip on the chip changer first.

Drag and Drop to Move the Chips

If you are dispensing using MANTIS RFID, the RFID icons and the Chip Lifetime Tracking indicators will automatically appear.

The Chip Lifetime Tracking represents each chip’s lifespan by tracking the chip cycles. Hover over the color bar in each input to see the chip lifespan status as a percentage.

  • Green (full) — Indicates that the chip is in good condition and ready for use.

  • Green — Indicates that the chip has 75% remaining cycles left. You can still use the chip for an optimal dispense.

  • Yellow — Indicates a higher usage of the chip lifespan. A chip with this status only has 50% to 25% remaining cycles left.

  • Red — The chip is about to reach its maximum lifespan. Replacing the chip is recommended to ensure continued dispense precision and accuracy.

  • Warning! — The chip has reached its maximum lifespan. Replacing the chip immediately is strongly recommended, as it presents higher risk of instrument damage if you continue using this chip. Any damaged parts caused by using an overused chip will not be covered under the warranty or service contract. For assistance, contact support@formulatrix.com.

  • Black (empty) — You cannot use the chip because it has 0% of cycle lifespan. Replace the chip with the new one to continue the dispense.

The Chip Lifetime Tracking Indicator in the Input Control Panel

Input Source Type You can configure your input source type in various ways.
Clicking the Input button enables you to change the input source type (tube or pipette tip), chip type (Low Volume, High Volume, Low Volume PFE, High Volume PFE, Low Volume 3PFE, High Volume 3PFE, or High Volume Continuous Flow), and configure the prime volume, wash volume, recovery volume, and pre-dispense volume. 

LV with the Tube Input Source Options

HV with the Pipette Input Source Options

Continuous Flow Options

  • Reagent Name — Contains the name of the reagent.

  • Chip Serial Number* — Contains the RFID chip's serial number.

  • Chip Type — Click the Chip Type list and select the desired chip type (Low Volume, High Volume, Low Volume PFE, High Volume PFE, Low Volume 3PFE, High Volume 3PFE, or Continuous Flow) that you want to use. The actual chip should match the chip type selected.

  • Input Source Type — Click the Input Source Type list and select the desired input source type (Tube or Pipette Tip) for your experiment. The recommended length for tubing is 450 mm. The HV and LV chips use a slightly different tubing. The HV tubing is green while the LV tubing is white. Please make sure you have matched the tube and chip accordingly.

    Note: This option will be automatically disabled and set to Tube by default if the Continuous Flow (CF) chip is assigned.

  • Prime Volume — Type the desired volume to fill your tubing and/or chip channels in the Prime Volume box. The MANTIS typically primes the chip only once at the beginning of the dispense to make sure all of the liquid channels are full. You will know the channels are full once reagent starts dispensing from the chip nozzle.

  • Prime Volume per Pulse**Type the desired volume for each pulse to fill your tubing and/or chip channels in the Prime Volume per Pulse box. The MANTIS typically primes the chip only once at the beginning of the dispense to make sure all of the liquid channels are full. You will know the channels are full once reagent starts dispensing from the chip nozzle.

  • Number of Priming Pulses** - Type the desired number of priming sequences that you want the MANTIS to perform. This allows for fine-tuning of priming behavior for difficult reagents.

  • Pre-Dispense VolumeType the desired reagent volume that you would like to dispense to waste before dispensing into your plate in the Pre-Dispense Volume box.

  • Recovery VolumeType the desired reagent volume that you would like to return to your source container (pipette tip or tube) in the Recovery Volume box.

  • Wash Step(s)Select the wash steps you want to execute.

    • 1 (Station 1) — This option enables MANTIS to wash the chip using only the wash liquid available in Wash Station 1 Source.

    • 2 (Station 1-2) — This option enables MANTIS to wash the chip using liquid in Wash Station 1 Source and rinse the chip using wash liquid in Wash Station 2 Source.

    • 3 (Station 1-2-1)This option enables MANTIS to do the 3-step wash. You must fill the Wash Station 1 Source with water and Wash Station 2 Source with wash solution. MANTIS will wash the chip using water in the Wash Station 1 Source, then wash solution in Wash Station 2 Source and rinse the chip using water in Wash Station 1 Source. For more information, see Sterilizing MANTIS Chips.

  • Wash Volume 1 (Station 1) — Type the desired wash volume to use during the chip cleaning process.

  • Wash Volume 2 (Station 2) — Type the desired wash volume to use during the chip cleaning process.

  • Wash Volume 3 (Station 1) — Type the desired wash volume to use during the chip cleaning process.

  • Recovery Duration** — Type the recovery duration in milliseconds in the Recovery Duration box.

  • Recovery Mode** — Select the type of recovery that you would like the MANTIS to do after a Continuous Flow dispense.

    • SingleSelect this option to enable one reagent recovery cycle. Use this mode for reagents with low to average viscosity.

    • DoubleSelect this option to enable two reagent recovery cycles. Use this mode for reagents with high viscosity.

    • PassiveSelect this option to enable MANTIS to run the reagent recovery cycle based on reagent volume, viscosity, and bottle position. This mode might take longer time than the other modes because it doesn’t use any pressure to recover the reagent to the bottle source.

Tip: If you are using a reagent with low viscosity, low surface tension, and/or a high possibility of creating foaming, such as water combined with detergent, use Passive mode. With Passive mode MANTIS will be able to prevent air bubbles during recovery, allowing more reagent to return to your source container in a single step. To ensure recovery of all of the reagent using Passive mode, it is recommended to set the Recovery Duration to more than 500 ms.

  • Delay Between Recovery**Type the delay time between two reagent recovery processes in milliseconds in the Delay Between Recovery box. This option is only available when you select the Double recovery mode.

  • Calibration Data** — Displays the calibration data for the current reagent, if the reagent has already been calibrated for CF dispense. To calibrate or recalibrate the reagent, click the Start New Calibration button. For more information, see Calibrating Continuous Flow Chips.

  • Pause for Reagent Refill*** — Enable this option to make the MANTIS pause during the dispense to refill the pipette tip.

  • Reagent Volume*** — Type the reagent volume that will cause MANTIS to pause for reagent refill. This option is available if you enable the Pause for Reagent Refill option. This option allows you to use a pipette tip with reagent volumes exceeding the tip capacity. For example, if you want to deliver 20 μL into every well of a 96 well plate while using 1 ml tip as input source (total 1920 μL) you can set the reagent volume to 950 μL. That will cause the MANTIS to pause for a top-up roughly half-way through the plate. Filling the tip with 10% extra sample is advised to prevent the reagent running out.

    Note: The reagent tracking works only during a dispense, and not between runs. For example, if you filled the tip with 1000 μL, then executed a dispense delivering 750 μL, after the dispense the reagent volume will remain set to 1000 μL in the software.

  • Restore Default ValuesClick the Restore Default Values button to restore the default values for all chip input settings.

  • Date of Manufacture* — Contains the chip’s manufacturing date.

  • Chip Dispense History tab* — Contains information about the chip usage and dispense process.


If you have multiple chips to which you wish to apply the same configuration parameters (like wash volumes or input source type), you do not need to configure each chip position individually. Click the Mass Update Input button to make changes to multiple chip stations at once.

Mass Update Input Options

Note: When you use mass update for RFID chips, MANTIS will automatically identify the chip type and the Chip Type option in the Mass Update Input window will be disabled.

Mass Update Input Window with the Assigned RFID Chips and Disabled Chip Type Option

You can also access, modify, and change the Input configuration through the Configuration All Input Settings button .

If your MANTIS is equipped with RFID, click the Scan All RFID Chips button to scan all RFID chips at once. For more information, See MANTIS Radio-Frequency Identification (RFID).


*Available for MANTIS with RFID.

**Available when you use the High Volume Continuous Flow chip.

***Available when you use the pipette tip input source.

Related Topics


MAOWH-V48R824

Unit 6, Level 13, Gate District, Gate Building

Dubai International Financial Centre

Dubai, United Arab Emirates

 

+1-781-788-0228

support@formulatrix.com

publications@formulatrix.com