LCP Experiments

Given the nature of an LCP experiment, the set up for an LCP dispense requires more hardware setup than other experiments. This tutorial walks you through hardware setup, provides two options for setting up the dispense in the software, running the LCP dispense, and finally attaching the LCP coverslip to the LCP base plate.

 

Set Up the NT8 Hardware

  1. Prepare the LCP syringe (for help with using the LCP mixer, see Mixing LCP Drops).

  2. Place the LV tip caddy in its station and the MRC 2 Well Crystallization Plate on microplate station 2. Each well of the MRC 2 plate should contain about 50 μl of solution.

  3. Prepare the LCP Plate:

  4. Attach the LCP syringe to the NT8’s head.

  5. Tip: For additional help, see Installing LCP Attachments.

  6. Use the lever to dispense a small amount of LCP sample from the syringe to prevent clogging and remove any air from the needle. Wipe the tip of the LCP syringe with a chem wipe.

  7. The sensors on the NT8 will detect the presence of the LCP syringe. A dialog box will open asking if you want to calibrate the syringe. Click Yes.

  8. Select the size of your LCP syringe and click OK. The NT8 will start calibrating the LCP syringe.

 

Option 1: Set Up the Software Using the LCP Template

An LCP dispense template is preinstalled with your software. You may use it if you plan to combine 800 nL well solution with 50 nL LCP drops. If you choose to use this template, you must also use the MRC 2 Well Crystallization plate as your well solution source.

 

To use this template:

  1. Click the Open Task List button .

  2. Open the Sample folder and select the LCP 50 nL Monolein And 800 nL Well Solution task list.

  3. Click Open.

  4. Run your experiment by clicking the Run Task List button .

 

Option 2: Manually Set Up the Software

  1. Drag the LCP Plate Adapter from the Plate Type library to Microplate Station 1.

  2. Find the plate type you used for your well solution and drag the plate to Microplate Station 2.

  3. Add the Pick Tip command to the Task List by double-clicking the LV Tip Caddy image.

  4. Click the Repeat button on the Task List to mark the beginning of tasks that will be repeated for every column.

  5. Add the Wash Tip command to the Task List by double-clicking the Wash Station image.

  6. Instruct the NT8 to aspirate well solution:

  7. Instruct the NT8 to dispense the LCP sample:

  8. Instruct the NT8 to dispense the well solution to the drop:

  9. Click End Repeat in the Task List to select it.

  10. Double-click the Waste Station image to add the Dispose Tip command to the Task List.

 

Run the LCP Dispense

 

Attach the LCP Coverslip

 