Configuring FRAP Bleaching Options

The FRAP Bleaching options are available on the FRAP Benchtop Rock Imager or Rock Imager 1000 with a FRAP imaging method option. On the Bleaching tab, you set options related to the laser, which is what bleaches the drop during FRAP imaging. To configure these settings, on the File menu, click FRAP Configurations, and click Bleaching.


Bleaching options

Bleach Options

Bleach Options provides two major settings, Pulse Laser for and Drop area to include in Random Bleach Position.

 
Option Description
Pulse Laser for The Pulse Laser for enables you to specify the amount of time the laser is focused on the drop to bleach the fluorophores.
Drop area to include in Random Bleach Position A bleach position is randomly chosen in the specified area. One or two different bleaching spots can be chosen; near the center of the drop or near the edge of the drop. Concentric rings define the boundary of the area in which a spot is chosen. If 25% area for Edge is specified, then a bleach spot will be performed on the outer 25% area of the drop, whereas the area for the center is defined from the center out.

 

Bleach Position Locator

The Bleach Position Locator is used to help the imager find areas within the LCP drop that are strong candidates for bleaching. Underexposed or overexposed areas within the drop will be ignored based on the values set with the sliders. Zero is black, and 255 is white. With the default settings, the imager will identify all areas within a drop that fall within the 80-255 pixel intensity range and use these areas as bleach spots.

 

Related Topics


RIC-V30R016

10 DeAngelo Drive

Bedford, MA 01730

USA

 

+1-781-788-0228

support@formulatrix.com

publications@formulatrix.com