TEMPEST Continuous Flow (CF) chip kit:
Item Name | Image |
---|---|
TEMPEST CF Cap — GL45 Bottle, PFE Harness with 2 Tubings (1/16 inch ID, 1/8 inch OD) | |
TEMPEST CF Chip | |
Nozzle Strip (0.6 mm) | |
GL45 Bottle, 250/500 mL size (Duran Pressure Plus or similar) |
Continuous Flow Chip Kit Assembly
You are running the TEMPEST software version 3.8 or newer.
You have connected the TEMPEST instrument to the laptop/PC.
Assemble the tubings, bottle cap, and GL45 bottle by following these steps:
Attach the CF bottle cap to the GL45 bottle. Pay attention to the gasket seal on the bottle cap and ensure it sits flat.
Gasket Seal Incorrect Alignment | Gasket Seal Correct Alignment |
Adjust the length and position of the tubings in the bottle by slightly loosening the three SHCS M3x8 screws using the 2.5 mm hex screwdriver until you can slide the tubes. Make sure you re-tighten the bolts after adjusting the tubing.
Loosening the Three SHCS M3x8 Screws
Turn on the TEMPEST instrument and open the software.
Go to the Chip Configuration settings by choosing from the following options:
Click the Setting buttonon the unassigned input in the Input Control Panel.
Click the Setting Button in the Input Control Panel
OR
Go to the Tools menu, and then select Options.
Select Options under the Tools Menu
In the Chip Configuration tab, click Move to Install Chip Position. At this stage, the TEMPEST head will automatically move forward to the chip installation position.
Click the Move to Install Chip Position Button
TEMPEST Head is Moving to the Installation Position
Note: A notification will appear to prompt you to recover any excess reagent before installing the CF chip. Click Yes to continue.
A Notification to Recover Any Excess Reagents
Install the CF chip to input position #1. Then, insert the two tubings into the input rail. Read the installing TEMPEST chip guide for more details.
Install the CF Chip to the Input Position #1
Insert the Two Tubings into the Input Rail
Insert the Air Pressure Source tubing from the back of the TEMPEST into the Push-to-Connect Fitting on the bottle cap.
Insert the Air Pressure Tubing into the Bottle Cap
Back to the Chip Configuration tab, input the installed CF chip Serial Number on the Chip 1 field. The Serial Number index must be TSCFxxx (for example, TSCF230106).
Enter the CF Chip Serial Number
Important: An errorindicator indicates that you have assigned the CF chip to an unallowable input position. The CF chip may only be installed in input position #1.
An Error Notification of Incorrect Chip Position
Click Save to save the configuration. See the Configuring the TEMPEST Chip section for more information.
After the CF chip has been properly installed, calibrate the chip from Continuous Flow Options under the Tools menu. Follow the Calibrating Continuous Flow Chips guide.
Continuous Flow Calibration Wizard
Congratulations! You are ready to design and run a dispense using a CF chip.
Continuous Flow Chip Dispensing Process