You are here: Continuous Flow : Setting Up TEMPEST Continuous Flow Chip for Dispense

Setting Up TEMPEST Continuous Flow Chip for Dispense

Required Parts:

Continuous Flow Chip Kit Assembly

Required Tool:

Prerequisites:

To set up the CF chip to TEMPEST:

  1. Assemble the tubings, bottle cap, and GL45 bottle by following these steps:

  2. Turn on the TEMPEST instrument and open the software.

  3. Go to the Chip Configuration settings by choosing from the following options:

  4. In the Chip Configuration tab, click Move to Install Chip Position. At this stage, the TEMPEST head will automatically move forward to the chip installation position.

    Click the Move to Install Chip Position Button

    TEMPEST Head is Moving to the Installation Position

    Note: A notification will appear to prompt you to recover any excess reagent before installing the CF chip. Click Yes to continue.

    A Notification to Recover Any Excess Reagents

  5. Install the CF chip to input position #1. Then, insert the two tubings into the input rail. Read the installing TEMPEST chip guide for more details.

    Install the CF Chip to the Input Position #1

    Insert the Two Tubings into the Input Rail

  6. Insert the Air Pressure Source tubing from the back of the TEMPEST into the Push-to-Connect Fitting on the bottle cap.

    Insert the Air Pressure Tubing into the Bottle Cap

  7. Back to the Chip Configuration tab, input the installed CF chip Serial Number on the Chip 1 field. The Serial Number index must be TSCFxxx (for example, TSCF230106).

    Enter the CF Chip Serial Number

    Important: An errorindicator indicates that you have assigned the CF chip to an unallowable input position. The CF chip may only be installed in input position #1.

    An Error Notification of Incorrect Chip Position

  8. Click Save to save the configuration. See the Configuring the TEMPEST Chip section for more information.

  9. After the CF chip has been properly installed, calibrate the chip from Continuous Flow Options under the Tools menu. Follow the Calibrating Continuous Flow Chips guide.

    Continuous Flow Calibration Wizard

  10. Congratulations! You are ready to design and run a dispense using a CF chip.