Release Notes¶
Version 4.13 Hotfix 1¶
Build Date: March 16, 2026¶
Changed:
The humidifier system configuration to match the current hardware.
Fixed:
[NT8-ROCK MAKER Integration] An error when aspirating seed in an experiment with the All Protein First then Well dispense and the Seed with Second seeding protocols.
[Configuration] The unapplied Delay configuration to the mixing process in Serial Dilution.
[Configuration] The unconfigured mixing speed for each liquid class.
[Plate Copy] The unapplied mixing cycle added after aspirating a liquid with a disabled liquid level sensor.
[Configuration] [Plate Copy] The incorrect default mixing speed value.
[NT8-ROCK MAKER Integration] The undisplayed screen in the Canvas tab after changing the Should Dispense Screen property in the Experiment Info tab to Yes.
The unapplied selected humidity level in the software to the hardware when connecting the software to the hardware.
[Task List] A software crash when selecting multiple columns for an aspirate or dispense task by clicking and dragging the cursor from any higher-number columns to column 1.
Version 4.13 BETA¶
Build Date: December 2, 2025¶
Added:
[NT8-ROCK MAKER Integration] Support for dispensing additive screens to a drop position.
[NT8-ROCK MAKER Integration] A feature to override the liquid class set in ROCK MAKER.
[Plate Type Library] The ability to support plates with more than 8 rows, such as 384 and 1536 well plates.
Support for dispensing of Serial Dilution experiments.
[NT8-ROCK MAKER Integration] The dispense converter from a drop to a well in plates without drops.
[Task List] The Group Row section to the Task Details dialog box to support plates with more than 8 rows.
Fixed:
[NT8-ROCK MAKER Integration] The incorrect single protein and high volume multiprotein block validation when running the No Seed seeding protocol in an experiment with a seed.
[NT8-ROCK MAKER Integration] The failure to dispense a well solution to a hanging drop experiment in a 24-well plate with All Protein First then Well dispense protocol.
[NT8-ROCK MAKER Integration] The failure to load two experiments containing the same protein but different dispensing targets.
[NT8-ROCK MAKER Integration] A software crash when selecting a plate in Plate Station 3.
[Plate Type Library] [Plate Type Editor] The NT8 head not moving to the correct well and drop positions after clicking Move To.
[Plate Type Library] [Plate Type Editor] Incorrect results of plate recalibration in an enabled LCP feature.
[NT8-ROCK MAKER Integration] [Experiment Info Tab] Unrounded seed decimal values when using an HV multiprotein block with enabled seed mixing.