Configuring Humidifiers

NT8 is equipped with leading-edge active humidity control that is powerful enough to moisten the air in the system anytime you need the NT8 up and running. What makes the NT8 stand out is that you can select any desired humidity target level. Both V3 and V4 hardware have up to three default value settings (75%, 80%, and 85%) for the humidity target you can choose from, ensuring that you will always be able to run aspiration and dispense tasks with the system at the perfect humidity level.

For example, if you set 85% as the target point, the system automatically regulates how many humidifiers are on to reach the predefined target. When the humidity level reaches the setpoint, the system automatically turns off the humidifier.

To configure your humidifiers in V4 hardware:

  1. In the Humidifier Control panel, click the Humidity Setting button humidity_setting_button.

    The Humidifier Control Panel

    The Humidifier Control Panel

  2. Determine the humidity level percentage in the Humidity Setting dialog box.

    The Humidity Setting Dialog Box

    The Humidity Setting Dialog Box

  3. Select Set.

  4. In the Setpoint checkboxes, select your preference of humidity level.

    Note

    The humidity level options in the Setpoint checkboxes will be based on each percentage you have set previously in the Humidity Setting dialog box.

    The Setpoint Checkboxes in the Humidifier Control Panel

    The Setpoint Checkboxes in the Humidifier Control Panel

To configure your humidifiers in V4 hardware:

  1. Go to the Humidifier Control panel and click the Humidity Setting button humidity_setting_button.

  2. Determine the humidity level percentage in the Humidifier Settings dialog box.

    The Humidifier Settings Dialog Box

    The Humidifier Settings Dialog Box

  3. Click Save.

  4. Open the Tools menu and select Options.

  5. In the Humidity Control item, select your preference of Humidity Level.

    Selecting Humidity Level in the Humidity Control

    Selecting Humidity Level in the Humidity Control

Related Topic