Using the Advanced Setup
The Advanced Setup feature allows you to define chip settings in a profile, including cleaning, mixing, and pressure control. You can customize these settings or create different setups for specific profiles. The Advanced Setup is accessible from the Editor page or during profile creation.
Overview
The Advanced Setup settings are divided into the following tabs:
General:
Chip Cleaning: Enable/disable chip cleaning before sample processing.
Note
New chips must be cleaned before use. Disable this setting only if the chips will be cleaned in a different profile.
Mixing: When enabled, aµtoPULSE periodically circulates the sample to minimize concentration gradients and prevent localized clogging. Specify the following settings to determine the sample circulation.
Mixing Every: Set the time interval between each mixing cycle.
Mixing Cycle: Define the length of each mixing cycle (in pumps).
![]()
General Tab
Cycle:
Diaphragm Pressure: Set the chip’s diaphragm pressure (in psi) to drive liquid flow at the membrane inlet (feed pressure).
Time Multiplier: Factor used to adjust the duration of each diaphragm stroke in the chip. Increasing the value makes each stroke longer.
Note
Thicker fluids need a higher multiplier to achieve optimal performance and recovery.
Return Valve Pressure: Set the return valve pressure (in psi) to control the membrane outlet pressure (retentate pressure). Specify the value for each kDa.
![]()
Cycle Tab
Creating a New Setup
Choose from the following options:
Open the General tab, define the chip’s cleaning and mixing settings.
Open the Cycle tab, specify the chip’s pressure-control settings.
Enter a new name and use SAVE AS to create a new setup.
Repeat steps 1–4 to create more setups.
Managing Existing Setups
To modify a setup:
To delete a setup:
