Labware Not Found

Issue: Insufficient labware for task execution.

Solving method:

  1. Turn OFF the scheduler via Cellmatic Suite.

  2. Register the labware.

  3. Open the Upper Storage window.

  4. Load the labwre to the Upper Storage according to their slots.

  5. Once done, close the Upper Storage window.

  6. Press the Pause button on Labware Storage. A Rover automatically scans the column accessed by the user. Once the scanning is complete, ensure all labware appears in the Cellmatic Suite.

  7. Delete the error in the Cellmatic Suite > Runtime Status > Failed Tasks.

  8. Turn on the scheduler via Cellmatic Suite.